Polyester (PET) is a hydrophilic membrane with good temperature resistance (up to 140 ° C) and a semi-translucent color.
Its chemical resistance is very high and in the case of biological sample filtration it is characterized by very low non-specific adsorption of proteins (<5 μg / cm2 of bovine serum albumin).
It is supplied in various pores of 0.22 μm and 0.45 μm; And in 25 and 47 mm in diameter.
Bubble point with water (minimium value)
0.2 µm 20.0 psi
0.45 µm 12.0 psi
Flow rate with wáter (ml/min/cm²) ∆p= 14.5 psi
0.2 µm 10
0.45 µm 33
Flow rate with air (L/min/cm²) ∆p= 10 psi
0.2 µm 3
0.45 µm 7.5
Thickness: 0.100 – 0.200 mm
Pore density:
0.2 µm 3 x 108 0.45 µm 1 x 108
Membrane material: Poliester
Diameter: 25 y 47 mm.
Maximum temperature in use ≤ 140ºC
Maximum work pressure: 8.7 psi
Extractables with water: < 1%
Protein absorption: 5 µg/cm2 Chemical compatibility: 4 – 8 pH
Reaction to water: hydrophilic
Sterilization: By autoclaving at 121 ° C or 134 ° C, dry heat at 180 ° C, γ-radiation, ethylene oxide