These filters with cellulose acetate membrane whose main properties are: low absorption and non specific adsorption of proteins, and their high flow rate. All these properties make them very valid for the filtration of aqueous samples and biological samples.
These syringe filters are available in a sterile format ( individual blister) and non sterile, in pores of 0.22 and 5.0 μm and in dimensions of 13, 25 and 30 mm of diameter.
Bubble point with water (minimum value)
0.2 µm 2.8 bar | 280 Kpa
0.45 µm 1.6 bar | 160 Kpa
0.8 µm 0.8 bar | 80 Kpa
1.2 µm 0.6 bar | 60 Kpa
5.0 µm 0.4 bar | 40 Kpa
Flow rate with water (ml/min /cm2)p=0.7 bar | 70 Kpa aprox.:
0.2 µm 10 ml/min/cm2
0.45 µm 30 ml/min/cm2
0.8 µm 45 ml/min/cm2
1.2 µm 100 ml/min/cm2
5.0 µm 200 ml/min/cm2
Thickness: 0.100 – 0.150 mm
Materials Membrane: Cellulose acetate
Housing: Polypropylene (PP)
Diameters: 13, 25, 30 mm.
Filtration area
13 mm diameter: 0.92 cm2
25 mm diameter : 2.98 cm2
30 mm diameter : 4,90 cm2
Maximum temperature of use ≤ 50ºC
Maximum pressure of work: 8.7 psi
Connectors
Input: Female luer lock
Output: Male luer slip
Hold-up volume
13 mm diameter: < 10 µl
25 mm diameter: < 30 µl
30 mm diameter: < 55 µl
Extractables with water < 0.2%
Chemical compatibility: 4 – 8 pH
Reaction to water: hydrophilic
Absorption: low absorption non-specific in proteins.